{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Tarrytown",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79496",
      "recalling_firm": "Siemens Healthcare Diagnostics, Inc.",
      "address_1": "511 Benedict Ave",
      "address_2": "N/A",
      "postal_code": "10591-5005",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Distributed in the following U.S. states: AL, CA, DE, GA, IA, IL, IN, KY, MA, MI, MS, MT, NC, ND, NJ, NM, NV, NY, OH, PA, RI, TN, TX, VA, WA, WV, WY and Puerto Rico. Distributed in the following OUS countries: Australia  Belgium¿¿¿¿¿¿¿¿  Canada  Chile  France¿¿¿¿¿¿¿¿¿  Germany¿¿¿¿¿¿¿¿  Greece¿¿¿¿¿¿¿¿¿  Israel  Italy¿¿¿¿¿¿¿¿¿¿  Japan  Latvia¿¿¿¿¿¿¿¿¿  Luxembourg¿¿¿¿¿  Netherlands¿¿¿¿  New Zealand  Norway¿¿¿¿¿¿¿¿¿  Poland¿¿¿¿¿¿¿¿¿  Portugal¿¿¿¿¿¿¿  Romania¿¿¿¿¿¿¿¿  Singapore  Slovakia¿¿¿¿¿¿¿  South Africa  Spain¿¿¿¿¿¿¿¿¿¿  Switzerland¿¿¿¿  U.A.E.¿¿¿¿¿¿¿¿¿  United¿Kingdom¿",
      "recall_number": "Z-2054-2018",
      "product_description": "ADVIA Centaur TSH3-Ultra (Thyroid Stimulating Hormone 3 Ultra), Catalog Numbers: 06491072 (100 tests), 06491080 (500 tests), 04862625 (2500 tests)",
      "product_quantity": "74,215",
      "reason_for_recall": "In certain scenarios the ADVIA Centaur CP, XP and XPT immunoassay analyzers may exhibit reduced dilution recovery or become inoperable when used with Multi-Diluent 15 and certain lots of BNP and TSH3RUL Assays.",
      "recall_initiation_date": "20171218",
      "center_classification_date": "20180601",
      "report_date": "20180613",
      "code_info": "Lot Numbers ending in: 298 and 301"
    }
  ]
}