  "results": [
    {
      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85147",
      "recalling_firm": "Philips North  America, LLC",
      "address_1": "3000 Minuteman Rd",
      "address_2": "N/A",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution.",
      "recall_number": "Z-2053-2020",
      "product_description": "Philips HeartStart XL Defibrillator/Monitor (Model number M4735A) - Product Usage: is for use in the hospital by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac life support, or defibrillation. It must be used by or on the order of a physician.",
      "product_quantity": "94,034 units",
      "reason_for_recall": "The rotary therapy selector switch may fail, resulting in unexpected device behavior including: 1) The device may not turn on; 2) The device may not perform the selected function;3) The device may deliver a shock with an energy level different than the setting selected by the user. In any of these scenarios, appropriate therapy delivery may be delayed.",
      "recall_initiation_date": "20200413",
      "center_classification_date": "20200520",
      "termination_date": "20210407",
      "report_date": "20200527",
      "code_info": "All units. Product has been discontinued."
    }
  ]
}