  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Bergen",
      "state": "N/A",
      "country": "Norway",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78466",
      "recalling_firm": "NordicNeuroLab AS",
      "address_1": "Mollendalsveien 65c",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "USA, Sweden, Spain, Russia, Hungary, India, Norway, Iceland, Italy, South Korea, United Kingdom",
      "recall_number": "Z-2050-2018",
      "product_description": "nordicBrainEx 2.0 provides analysis and visualization capabilities of dynamic MRI data of the brain, presenting the derived properties and parameters in a clinically useful context.",
      "product_quantity": "22 licenses",
      "reason_for_recall": "An error was discovered in the interpretation of certain DICOM header tags that may lead to incorrect orientation labeling, and thus and indirect left-right, up-down or anterior-posterior flipping of images.",
      "recall_initiation_date": "20140917",
      "center_classification_date": "20180601",
      "termination_date": "20210324",
      "report_date": "20180613",
      "code_info": "version 2.1 and older"
    }
  ]
}