  "results": [
    {
      "status": "Terminated",
      "city": "Indianapolis",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74191",
      "recalling_firm": "Nico Corp.",
      "address_1": "250 E 96th St Ste 125",
      "address_2": "N/A",
      "postal_code": "46240-3872",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Distribution to the states of : OR, IL, IA, NC, TX, CA, GA, OH, WI, AR, MI, MD, IN.",
      "recall_number": "Z-2045-2016",
      "product_description": "Myriad handpiece  Tissue morcellator",
      "product_quantity": "63 units",
      "reason_for_recall": "This recall has been initiated due to a defective component within the handpiece which will  cause the handpiece to fail. The failure will result in the inability to cut tissue with the  handpiece, rendering it non-functional. Should this failure occur, a replacement Myriad  handpiece will be required. If a replacement handpiece is not available, alternate instrumentation for tissue removal will be necessary.",
      "recall_initiation_date": "20140224",
      "center_classification_date": "20160621",
      "termination_date": "20160628",
      "report_date": "20160629",
      "code_info": "13ga, 13cm Myriad Handpiece  Part number: NN-8004  Lots 1444664 and 1508994"
    }
  ]
}