  "results": [
    {
      "status": "Terminated",
      "city": "San Diego",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71404",
      "recalling_firm": "Volcano Corp",
      "address_1": "3721 Valley Centre Dr Ste 500",
      "address_2": "N/A",
      "postal_code": "92130-3328",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide and the countries of Belgium, United Kingdom, Denmark, Netherlands, Sweden, Iceland, France, Saudi Arabia, Australia, Philippines and Thailand.",
      "recall_number": "Z-2045-2015",
      "product_description": "ComboMap Intravascular Flow and  Pressure System;  Product/Part number: Combo Map - 802745-001.       Intravascular Imaging and Pressure System for evaluation of blood flow, perfusion and vascular structures.",
      "product_quantity": "227 total units - all models",
      "reason_for_recall": "A limitation in interoperability between the Philips Xper Flex Cardia Physiomonitoring System and Volcano's sS/Si, Core, and ComboMap systems. This limitation may impact the calculation of Fraction Flow Reserve (FFR) within the Volcano System when the system is connected to the Xper Flex Cardia system via its aortic pressure auxiliary output.",
      "recall_initiation_date": "20130923",
      "center_classification_date": "20150708",
      "termination_date": "20150709",
      "report_date": "20150715",
      "code_info": "Product/Part numbers: Combo Map - 802745-001.    All software versions"
    }
  ]
}