      "status": "Terminated",
      "city": "Bergen",
      "state": "N/A",
      "country": "Norway",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "78463",
      "recalling_firm": "NordicNeuroLab AS",
      "address_1": "Mollendalsveien 65c",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "Worldwide Distribution -- USA, Australia, Denmark, China, France, Japan, Malaysia, Netherlands, USA, Poland, Sweden, Spain, Russia, India, Norway, Singapore, South Korea, United Kingdom, and Slovenia.",
      "recall_number": "Z-2044-2018",
      "product_description": "nordicICE 2.3.14    Image processing software package used by trained professionals,  including physicians and medical technicians.",
      "product_quantity": "97 licenses",
      "reason_for_recall": "An error was discovered in the interpretation of certain DICOM header tags that may lead to incorrect orientation labeling, and thus and indirect left-right, up-down or anterior-posterior flipping of images.",
      "recall_initiation_date": "20140912",
      "center_classification_date": "20180531",
      "termination_date": "20201006",
      "report_date": "20180606",
      "code_info": "Version 2.3.14"
    }
  ]
}