  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Cleveland",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68643",
      "recalling_firm": "Philips Medical Systems (Cleveland) Inc",
      "address_1": "595 Miner Rd",
      "address_2": "",
      "postal_code": "44143-2131",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed in the states of CA, CO, IL, OK, OR and WI.",
      "recall_number": "Z-2037-2014",
      "product_description": "GEMINI TF PET/CT16: 882470, GEMINI TF PET/CT 64: 882471, GEMINI TF Big Bore: 882476.    A diagnostic imaging system for fixed or mobile installations that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems.",
      "product_quantity": "33",
      "reason_for_recall": "When beginning a procedure to start the gantry and CT Host, a Close Estop\u001d message may appear to the operator, but disappears before action can be taken. If this occurs it results in an inability to close E-Stop and the system is not operational for clinical use. This problem occurs randomly.",
      "recall_initiation_date": "20140618",
      "center_classification_date": "20140710",
      "termination_date": "20150916",
      "report_date": "20140716",
      "code_info": "GEMINI TF PET/CT16: 882470  7019,7039,7099,7588,7593,7598,7599,7600,7604,7605,7596M    GEMINI TF PET/CT 64: 882471  7027,7044,7146,7153,7178,7182,7188,7578,7579,7585,7587,7589,7592,7594,7595,7602,  7603, 7606    GEMINI TF Big Bore: 882476  9019,9227,9232,9233",
      "more_code_info": ""
    }
  ]
}