  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Twinsburg",
      "address_1": "1959 Summit Commerce Park",
      "reason_for_recall": "Hitachi discovered a software error in the Reconstruction Status function that creates a potential risk that a patient study will not be fully reconstructed. When multiple reconstructions are queued for processing, if a queue is suspended and the task order changed, the second queue may not reconstruct the full range of slices. This may result in missing diagnostic data.",
      "address_2": "",
      "product_quantity": "Six (6) units",
      "code_info": "Serial #'s S5001 - S5011",
      "center_classification_date": "20120719",
      "distribution_pattern": "US Nationwide Distribution to the following states of: MD, MI, MX, ND, NY, OH and TX.",
      "state": "OH",
      "product_description": "Hitachi Scenaria CT System Software    Product Usage:  The Scenaria system is indicated to acquire axial volumes of the whole body including the head. The images can be acquired in either axial, helical, gated or dynamic modes. The volume datasets acquired by the Scenaria can be post processed by the Scenaria to provide additional information. Post processing capabilities included in the Scenaria software include CT angiography (CTA), Multi-planar reconstruction (MPR) and volume rendering.",
      "report_date": "20120725",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Hitachi Medical Systems America Inc",
      "recall_number": "Z-2037-2012",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "product_type": "Devices",
      "event_id": "62372",
      "termination_date": "20130219",
      "more_code_info": "",
      "recall_initiation_date": "20120504",
      "postal_code": "44087",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}