      "country": "United States",
      "city": "Research Triangle Park",
      "address_1": "2917 Weck Dr.",
      "reason_for_recall": "Speaker Failure. If the speaker fails and an event occurs which would generate an audio alarm, the alarm will not sound and could cause a potential delay in patient treatment.",
      "address_2": "",
      "product_quantity": "14,794 each",
      "code_info": "Catalog #425-00, 425-10 and 425-30.",
      "center_classification_date": "20120719",
      "distribution_pattern": "Worldwide Distribution - U.S. (nationwide) and the countries of  Argentina, Australia, Brazil, Canada, Chile, China, Europe, India, Japan, Mexico, South Korea, Taiwan and UK.",
      "state": "NC",
      "product_description": "Hudson RCI Concha Therm Neptune Humidifier, Rx Only,     Product Usage:  Usage:  Device that is intended to add moisture to, and sometime to warm, the breathing gases for administration to a patient.",
      "report_date": "20120725",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Teleflex Medical",
      "recall_number": "Z-2036-2012",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "62284",
      "termination_date": "20140128",
      "more_code_info": "",
      "recall_initiation_date": "20120612",
      "postal_code": "27709",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}