      "status": "Terminated",
      "city": "Snoqualmie",
      "state": "WA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83220",
      "recalling_firm": "Spacelabs Healthcare, Inc.",
      "address_1": "35301 Se Center St",
      "address_2": "N/A",
      "postal_code": "98065-9216",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distribution was nationwide.  There was also government and military distribution.    Foreign distribution was made to Argentina, Bahrain, Bolivia, Brazil, Canada, Cayman Islands, Chile, China, Costa Rica, Czech Republic, France, Hong Kong, India, Indonesia, Italy, Kuwait, Malaysia, Mexico, Netherlands, Pakistan, Palestine, Panama, Philippines, Poland, Qatar, Romania, Saudi Arabia, Spain, Switzerland, Taiwan, Trinidad and Tobago, Tunisia, Turkey, UAE, United Kingdom, and Viet Nam.",
      "recall_number": "Z-2035-2019",
      "product_description": "Spacelabs Healthcare Smart Disclosure System, Model 92810, a component of the Intesys Clinical Suite (ICS) G2.",
      "product_quantity": "295 customer sites have the affected software versions",
      "reason_for_recall": "Several reports were received that patient records were printed with the correct patient demographics but containing another patient's waveforms.",
      "recall_initiation_date": "20190128",
      "center_classification_date": "20190722",
      "termination_date": "20220224",
      "report_date": "20190731",
      "code_info": "All affected Smart Disclosure installations with software versions 5.00, 5.01, 5.02, or 5.03, UDI 10841522123580."
    }
  ]
}