      "status": "Terminated",
      "city": "Andover",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "79982",
      "recalling_firm": "Philips Electronics North  America Corporation",
      "address_1": "3000 Minuteman Rd",
      "address_2": "N/A",
      "postal_code": "01810-1032",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution to accounts in OH and NC. Foreign distribution to Lebanon.",
      "recall_number": "Z-2035-2018",
      "product_description": "HeartStart MRx Monitor/Defibrillator, Model No. M3535A (861288), M3536A (861289)    Product Automated external defibrillators (non-wearable)    Product Usage:  The HeartStart MRx is for use for the termination of ventricular tachycardia and ventricular fibrillation. The device is for use by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced cardiac support, or defibrillation. It must be used by or on the order of a physician.",
      "product_quantity": "5",
      "reason_for_recall": "The MRx monitor/defibrillators could fail to charge because the therapy printed circuit board may have been loaded with an incorrect electronic component.  Failure to charge could potentially cause therapy to be interrupted or delayed.",
      "recall_initiation_date": "20180503",
      "center_classification_date": "20180530",
      "termination_date": "20191017",
      "report_date": "20180606",
      "code_info": "US00588464, US00588657, US00589878, US00591080, US00593861"
    }
  ]
}