  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Arlington",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "80113",
      "recalling_firm": "Joerns Healthcare",
      "address_1": "2100 Design Rd",
      "address_2": "N/A",
      "postal_code": "76014-4593",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide in the states of UK, Canada, Holland",
      "recall_number": "Z-2034-2018",
      "product_description": "ProCair Mattress:  (a) PROCAIR PLUS 42X80, MATERIAL NUMBER 222-4280  (b) PROCAIR PLUS 36X80, MATERIAL NUMBER 222-3680  (c) PROCAIR PLUS 84\" MATTRESS, MATERIAL NUMBER 222-6384    Product Usage:  Non powered flotation therapy mattress. The device is intended to treat or prevent decubitus ulcers (bed sores).",
      "product_quantity": "46 units",
      "reason_for_recall": "Joerns Healthcare has identified a potential issue with some of the P.R.O. Matt and ProCair Mattresses.  For the identified serial number range of P.R.O. Matt and ProCair Mattresses incorrect connectors (part #s 206-0401 & 206-0405) were provided by a Joerns supplier without the required internal shutoff valves, these shutoff valves prevent air from escaping from the mattress air cells.",
      "recall_initiation_date": "20180503",
      "center_classification_date": "20180530",
      "termination_date": "20230928",
      "report_date": "20180606",
      "code_info": "(a) PROCAIR PLUS 42X80, MATERIAL NUMBER 222-4280  Serial Numbers:  1000054866  1000054860  1000051368  1000054859  1000054855  1000054857    (b) PROCAIR PLUS 36X80, MATERIAL NUMBER 222-3680  Serial Numbers:  1000053319  1000053312  1000052886  1000053325  1000053323  1000052420  1000052259  1000052426  1000052425  1000052432  1000052891  1000052458  1000052429  1000052256  1000052888  1000053990  1000051418  1000051419  1000052457  1000052252  1000051411  1000055556  1000052460  1000054129  1000053979  1000049129  1000051412  1000052424  1000052455  1000054117  1000053982  1000053983  1000054122  1000054121    (c) PROCAIR PLUS 84\" MATTRESS, MATERIAL NUMBER 222-6384  Serial Numbers:  1000055501  1000049132  1000055498  1000057147  1000055500  1000056219"
    }
  ]
}