  "results": [
    {
      "country": "United States",
      "city": "Allendale",
      "address_1": "2 Pearl Ct",
      "reason_for_recall": "Beginning in February 2011, Stryker Spine began receiving reports related to the shaft of the AVS Navigator trial failing at the junction of the trial head.",
      "address_2": "",
      "product_quantity": "273 units",
      "code_info": "Class I; 510(k) Exempt  Description:  AVS¿ Navigator Trials. Size 6mm -15mm.  Catalog Number:  48392206,48392207,48392208,48392209,48392210, 48392211,48392212,48392213,48392214,48392215.  Lot Code:  104864,104866, 104867, 104868, 104869, 104870, 10E331, 104871, 104872, 104873, 104874, 104875.",
      "center_classification_date": "20120719",
      "distribution_pattern": "Nationwide.",
      "state": "NJ",
      "product_description": "Stryker AVS Navigator Trials 6mm-15mm  Manufactured by:  Stryker Spine SAS  Z.1 Marticot-33610-Cestas -France  +33 (0) 5.57.97.06.30  http;//www.stryker.com    Distributed in the USA by   Stryker Spine  2 Pearl Court. Allendale  NJ 07401-1677 USA  +1-201-780-8000  Cestas France 33610.    Intended Use: The trials are designed and intended to measure the height of the disc space intra-operatively.",
      "report_date": "20120725",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Stryker Spine",
      "recall_number": "Z-2033-2012",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "61723",
      "termination_date": "20130729",
      "more_code_info": "",
      "recall_initiation_date": "20120113",
      "postal_code": "07401-1611",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}