  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Mercedita",
      "state": "PR",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68547",
      "recalling_firm": "Zimmer Manufacturing B.V.",
      "address_1": "Turpeaux Industrial Park Rd 1 km 123.4",
      "address_2": "",
      "postal_code": "00715",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - USA (nationwide) and the countries of  India, Japan, China, Australia and Canada.",
      "recall_number": "Z-2032-2014",
      "product_description": "Zimmer  EDI  PERIARTICULAR PLATING SYSTEM  CANCELLOUS BONE SCREW, FULLY THREADED, HEX HEAD 4MM DIAMETER, 14 MM LENGTH, and EDI 4.0 MM CANCELLOUS SCREW, FULLY THREADED, 20 MM LENGTH.   Temporary internal fixation devices, designed to stabilize fractures during the normal healing process.",
      "product_quantity": "496",
      "reason_for_recall": "Two lots of screws were commingled.  Screws with etching and machining for lot number 62629432 were packaged in bags labeled for lot number 62628759; and screws with etching and machining for lot number 62628759 were packaged in bags labeled for lot number 62629432.  The affected product was distributed between 3/6/14 and 4/21/14.",
      "recall_initiation_date": "20140612",
      "center_classification_date": "20140710",
      "termination_date": "20150624",
      "report_date": "20140716",
      "code_info": "Lot No. 62629432  Lot No. 62628759",
      "more_code_info": ""
    }
  ]
}