  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Sparks",
      "address_1": "BD Diagnostic Systems",
      "reason_for_recall": "Potential for increase level of artifacts to appear as large, spherical and easy to distinguish or small & more difficult to distinguish from gram positive cocci.",
      "address_2": "7 Loveton Circle",
      "product_quantity": "878 cartons/ 3512 bottles",
      "code_info": "Lots 1325324 Exp May 31,2013 and 2011746, Exp July 31,2013",
      "center_classification_date": "20120718",
      "distribution_pattern": "Nationwide Distribution.",
      "state": "MD",
      "product_description": "Gram Crystal Violet 250 mL, catalog #212525, in cartons of 4 bottles.     Gram Stain Kits and Reagents are used to stain microorganisms from cultures or specimens by the differential Gram method.",
      "report_date": "20120725",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Becton Dickinson & Co.",
      "recall_number": "Z-2032-2012",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "61769",
      "termination_date": "20121210",
      "more_code_info": "",
      "recall_initiation_date": "20120412",
      "postal_code": "21152-0999",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}