{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98664",
      "recalling_firm": "Medline Industries, LP",
      "address_1": "3 Lakes Dr",
      "address_2": "",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution.",
      "recall_number": "Z-2031-2026",
      "product_description": "Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows:  GENERAL KIT, Medline Kit Number/SKU DYNJ905664I",
      "product_quantity": "380 kits",
      "reason_for_recall": "During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.",
      "recall_initiation_date": "20260319",
      "center_classification_date": "20260430",
      "report_date": "20260506",
      "code_info": "UDI/DI each 10195327487546, UDI/DI case 40195327487547, Lot Numbers:  24KME456, 24HMJ900.",
      "more_code_info": ""
    }
  ]
}