  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83203",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-2031-2019",
      "product_description": "Artis zee biplane, Model Number 10094141",
      "product_quantity": "268",
      "reason_for_recall": "A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.",
      "recall_initiation_date": "20190612",
      "center_classification_date": "20190722",
      "termination_date": "20200918",
      "report_date": "20190731",
      "code_info": "Serial Numbers: 153131  153229  153172  153315  153317  153992  153335  153695  153101  153102  153717  153779  153280  153648  153674  153166  153700  154096  153834  153194  153169  153803  153110  153890  154157  153199  153887  153269  154074  154012  153191  153252  153732  153239  154090  153227  153889  153655  153234  153236  153285  154686  153273  153251  153254  153344  154095  153342  154093  153171  153197  153916  153831  153256  153154  154086  153139  153882  153311  153211  153118  153987  153740  153863  153772  154079  153965  153707  154098  154073  153263  153276  154076  153758  131013  153257  154156  153351  153112  153155  153245  153318  153720  153998  153271  153857  153226  153939  153951  153290  153678  153679  153277  153125  153128  153134  153135  153743  153967  153946  153145  153866  153868  153829  153148  131021  153878  153906  153959  153709  153767  153217  153639  153119  153164  153304  153869  153719  153232  131005  153770  154020  153930  153728  153355  154058  153811  153225  154155  153883  153875  153289  153664  153963  153824  154034  154018  153356  153701  153113  153235  153347  153708  153338  154066  153259  153314  154014  153988  154048  153716  153215  153922  153870  153107  153178  153249  153706  153860  153264  153157  153903  153908  153819  154087  153914  154085  153156  153697  131000  153147  153790  153928  153864  153159  153756  153808  154052  153797  153634  154028  154084  153201  153865  153640  153714  153736  153832  153913  153348  153220  153947  153657  153873  153109  153150  153299  153757  153898  154075  154064  153111  153618  153689  154044  153847  154056  154035  153999  153124  153682  153120  153122  153329  153173  153763  153686  154100  154082  153617  153327  153180  153818  154007  153323  154055  154067  154088  131020  153762  153270  153862  153670  153607  153760  154094  153265  153694  153886  153332  153141  153195  153920  153343  153975  153993  153325  153631  153160  153186  153729  154089  153354  153684  153799  154041  153742  153952  154077  153214  153722  153206  153255  154097  153115  153744  153827  153671"
    }
  ]
}