  "results": [
    {
      "status": "Terminated",
      "city": "Knoxville",
      "state": "TN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74192",
      "recalling_firm": "Siemens Medical Solutions USA Inc.",
      "address_1": "810 Innovation Dr",
      "address_2": "N/A",
      "postal_code": "37932-2562",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide",
      "recall_number": "Z-2031-2016",
      "product_description": "Biograph mCT-S(40) 4R, MATERIAL NUMBER 10248671    The Siemens Biograph TruePoint systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.",
      "product_quantity": "100 units",
      "reason_for_recall": "Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.",
      "recall_initiation_date": "20160422",
      "center_classification_date": "20160621",
      "termination_date": "20170508",
      "report_date": "20160629",
      "code_info": "Serial Numbers:  1024,1001,1002,1003,1005,1006,1007,1008,1009,1011,1012,1013,1004,1014,1017,1018,1019,1020,1022,1010,1023,11034,1016,1015,1025,11026,11029,11028,11030,11031,11035,11032,11033,11027,11036,11037,11039,11038,11040,11043,11044,11045,11046,11083,11047,11041,11042,11049,11051,11050,11053,11054,11055,11056,11057,11058,11059,11076,11060,11061,11062,11063,11064,11052,11065,11048,11074,11069,11067,11068,11070,11072,11075,11073,11066,11077,11078,11071,11080,11081,11082,11084,21085,21086,21087,21088,21089,21090,21091,31092,21093,31094,31095,31097,31096,31099,91001,31100,1021,31098"
    }
  ]
}