{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-05-06",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Northfield",
      "state": "IL",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98664",
      "recalling_firm": "Medline Industries, LP",
      "address_1": "3 Lakes Dr",
      "address_2": "",
      "postal_code": "60093-2753",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution.",
      "recall_number": "Z-2030-2026",
      "product_description": "Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows:    1. MICRODISECTOMY PACK, Medline Kit Number/SKU DYNJ80585B;   2. MICRODISECTOMY PACK, Medline Kit Number/SKU DYNJ80585C;   3, MICRODISECTOMY PACK, Medline Kit Number/SKU DYNJ80585D.",
      "product_quantity": "924 kits",
      "reason_for_recall": "During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.",
      "recall_initiation_date": "20260319",
      "center_classification_date": "20260430",
      "report_date": "20260506",
      "code_info": "Medline Kit Number/SKU DYNJ80585B:  UDI/DI each 10195327211158, UDI/DI case 40195327211159, Lot Number: 22GBV101;   Medline Kit Number/SKU DYNJ80585B:  UDI/DI each 10195327211158, UDI/DI case 40195327211159, Lot Number: 24IBG473;   Medline Kit Number/SKU DYNJ80585C:  UDI/DI each 10198459154751, UDI/DI case 40198459154752, Lot Number: 25EBT656;   Medline Kit Number/SKU DYNJ80585C:  UDI/DI each 10198459154751, UDI/DI case 40198459154752, Lot Number: 25DBN421;   Medline Kit Number/SKU DYNJ80585C:  UDI/DI each 10198459154751, UDI/DI case 40198459154752, Lot Number: 25CBN640;   Medline Kit Number/SKU DYNJ80585C:  UDI/DI each 10198459154751, UDI/DI case 40198459154752, Lot Number: 25BBL172;   Medline Kit Number/SKU DYNJ80585C:  UDI/DI each 10198459154751, UDI/DI case 40198459154752, Lot Number: 25BBD651;   Medline Kit Number/SKU DYNJ80585C:  UDI/DI each 10198459154751, UDI/DI case 40198459154752, Lot Number: 25ABA455;   Medline Kit Number/SKU DYNJ80585C:  UDI/DI each 10198459154751, UDI/DI case 40198459154752, Lot Number: 24KBO858;   Medline Kit Number/SKU DYNJ80585C:  UDI/DI each 10198459154751, UDI/DI case 40198459154752, Lot Number: 24JBH693;   Medline Kit Number/SKU DYNJ80585D:  UDI/DI each 10198459389337, UDI/DI case 40198459389338, Lot Number: 26ABH754;   Medline Kit Number/SKU DYNJ80585D:  UDI/DI each 10198459389337, UDI/DI case 40198459389338, Lot Number: 25LBA242;   Medline Kit Number/SKU DYNJ80585D:  UDI/DI each 10198459389337, UDI/DI case 40198459389338, Lot Number: 25HBT362;   Medline Kit Number/SKU DYNJ80585D:  UDI/DI each 10198459389337, UDI/DI case 40198459389338, Lot Number: 25GBL430;   Medline Kit Number/SKU DYNJ80585D:  UDI/DI each 10198459389337, UDI/DI case 40198459389338, Lot Number: 25FBT402.",
      "more_code_info": ""
    }
  ]
}