{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83203",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-2028-2019",
      "product_description": "Artis zee floor, Model Number 10094135",
      "product_quantity": "170",
      "reason_for_recall": "A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.",
      "recall_initiation_date": "20190612",
      "center_classification_date": "20190722",
      "termination_date": "20200918",
      "report_date": "20190731",
      "code_info": "Serial Numbers: 135944  135746  135120  135322  135757  136513  136562  135914  135102  136042  135927  135986  135935  135953  135841  135200  135201  135279  135169  135879  135229  135157  136113  135990  136077  136103  135862  135235  135117  135358  135920  136168  135148  135366  136568  135843  135111  135112  135779  136224  135137  136607  136579  135309  124007  124015  136280  136286  135109  135413  135144  135145  136362  135404  136318  136234  135151  135421  135264  135199  135760  136500  136515  135217  136031  136032  135124  135125  136084  136026  136479  135179  135394  135285  136338  136196  135245  135420  136384  136387  136488  135770  136550  136256  136408  136450  136508  136547  135788  136242  136383  136517  135406  136043  136586  135129  135416  135909  135222  136393  136520  135258  135260  135755  136548  135143  135114  135133  136350  135353  135253  135375  135772  136483  136555  135190  135881  136614  137384  136567  135714  135922  136314  136190  136197  136254  135813  135791  136561  135302  135152  136505  135182  124006  135868  135793  135132  135723  136126  135847  135281  135373  135196  136128  136147  136557  136558  136391  136503  136143  135352  135388  135384  136582  136345  136570  135954  136459  136518  135771  136290  136609  135233  136594  135203  136137  135286  135397  136278  135424"
    }
  ]
}