{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Louisville",
      "state": "KY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92448",
      "recalling_firm": "DRE Medical Group Inc",
      "address_1": "2601 Stanley Gault Pkwy Ste 101",
      "address_2": "N/A",
      "postal_code": "40223",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US: CA, FL",
      "recall_number": "Z-2025-2023",
      "product_description": "Philips Respironics V60 Ventilator-Intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Model Number: R1053618 and R1076709.  These are retrofitted device.",
      "product_quantity": "22 units",
      "reason_for_recall": "Devices were imported with Japanese Language and NIST Oxygen Gas Fittings and retrofitted, may pose a serious health risk as the gas pathway of these devices may be contaminated.",
      "recall_initiation_date": "20230515",
      "center_classification_date": "20230628",
      "report_date": "20230705",
      "code_info": "USI-DI: 00884838009868 Serial Numbers: Serial Number 100043954 100050078 100052760 100052848 100054619 100054626 100055277 100055557 100058549 100058566 100058837 100058995 100060228 100065373 100069520 100070487 100071122 100072691 100078379 100080232 100080844 201003462",
      "more_code_info": ""
    }
  ]
}