  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Elyria",
      "state": "OH",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68399",
      "recalling_firm": "Invacare Corporation",
      "address_1": "1200 Taylor St",
      "address_2": "",
      "postal_code": "44035-6248",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed in PA and Canada.",
      "recall_number": "Z-2023-2014",
      "product_description": "Aquatec Fixed Backrest - accessory for Mobile Shower and Toilet commode Aquatec Ocean, Ocean VIP and Ocean E-VIP family.",
      "product_quantity": "465 units",
      "reason_for_recall": "The Aquatec¿ Fixed Backrest withstands a load of only 250 newtons at uneven load. This may result in the Fixed Backrest becoming detached from the frame.",
      "recall_initiation_date": "20140625",
      "center_classification_date": "20140709",
      "termination_date": "20150415",
      "report_date": "20140716",
      "code_info": "Mobile Shower and Toilet commode Aquatec Ocean (Model # 1534327), Ocean VIP (Model # 1470713 and 1470793) and Ocean E-VIP (Model # 1525707).",
      "more_code_info": ""
    }
  ]
}