{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83203",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-2022-2019",
      "product_description": "AXIOM Artis dBA, Model Number 7555357",
      "product_quantity": "85",
      "reason_for_recall": "A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.",
      "recall_initiation_date": "20190612",
      "center_classification_date": "20190722",
      "termination_date": "20200918",
      "report_date": "20190731",
      "code_info": "Serial Numbers: 53181  53023  53167  53104  53067  53093  53162  53706  53058  53098  53253  53016  53188  53209  53223  53700  53017  53114  53063  53219  53033  53142  53116  53002  53086  53195  53165  53076  53064  53124  53176  53078  53111  53227  53202  53133  53140  53047  53710  53190  53180  53128  53185  53170  53187  53019  53141  53084  53074  53225  53215  53189  53056  53166  53149  53020  53096  53009  53147  53070  53049  53261  53122  53065  53031  53085  53026  53222  53090  53075  53256  53045  53057  53103  53182  53012  53022  53175  53194  53102  53109  53260  53043  53160  53117"
    }
  ]
}