  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Knoxville",
      "state": "TN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74192",
      "recalling_firm": "Siemens Medical Solutions USA Inc.",
      "address_1": "810 Innovation Dr",
      "address_2": "N/A",
      "postal_code": "37932-2562",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide",
      "recall_number": "Z-2022-2016",
      "product_description": "BIOGRAPH mCT Flow 40-3R, MATERIAL NUMBER 10529158    The Siemens Biograph TruePoint systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.",
      "product_quantity": "3 units",
      "reason_for_recall": "Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.",
      "recall_initiation_date": "20160422",
      "center_classification_date": "20160621",
      "termination_date": "20170508",
      "report_date": "20160629",
      "code_info": "Serial Numbers:  11001,11002,11003"
    }
  ]
}