  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83203",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-2021-2019",
      "product_description": "AXIOM Artis dTC, Model Number 7413078",
      "product_quantity": "98",
      "reason_for_recall": "A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.",
      "recall_initiation_date": "20190612",
      "center_classification_date": "20190722",
      "termination_date": "20200918",
      "report_date": "20190731",
      "code_info": "Serial Numbers: 46022  46023  46154  46107  46263  46091  46203  46100  46025  46034  46042  46063  46036  46245  46062  46198  46169  46272  46117  46051  46137  46157  46215  46216  46219  46220  46221  46108  46040  46124  46126  46225  46226  46120  46152  46223  46130  46241  46213  46128  46205  46231  46098  46133  46271  46060  46085  46201  46182  46122  46065  46143  46144  46186  46097  46111  46067  46138  46269  46054  46125  46194  46190  46168  46180  46044  46127  46204  46112  46115  46118  46150  46164  46103  46113  46270  46178  46047  46106  46156  46202  46167  46163  46045  46158  46210  46179  46256  46192  46176  46170  46171  46174  46101  46139  46061  46035  46239"
    }
  ]
}