  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Leesburg",
      "state": "VA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85491",
      "recalling_firm": "K2M, Inc",
      "address_1": "600 Hope Pkwy SE",
      "address_2": "N/A",
      "postal_code": "20175-4428",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide distribution including in the states of Colorado, Michigan, Texas, California, New York, Florida, Maryland, Oregon, Kansas, Montana, Missouri and the country of Europe.",
      "recall_number": "Z-2020-2020",
      "product_description": "K2M, ALEUTIAN Interbody Spacer Systems, EN AN Lordotic Spreader, Stainless Steel, Non-Sterile, Qty 1, Catalog # 5203-90149, UDI # 10888857108622, Size 28x4mm, 6  Product Usage: Used to simulate the height of the cage and expand the spine disc space to determine the optimal implant size base on patient anatomy.",
      "product_quantity": "N/A",
      "reason_for_recall": "The Aleutian Anatomically Narrow Lordotic Spreader was manufactured as two pieces welded together which has been associated with the spreader tip breaking off during the procedure.",
      "recall_initiation_date": "20200413",
      "center_classification_date": "20200515",
      "termination_date": "20211115",
      "report_date": "20200527",
      "code_info": "Lot # FKXJ"
    }
  ]
}