  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83203",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-2020-2019",
      "product_description": "AXIOM Artis dFC, Model Number 7412807",
      "product_quantity": "148",
      "reason_for_recall": "A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.",
      "recall_initiation_date": "20190612",
      "center_classification_date": "20190722",
      "termination_date": "20200918",
      "report_date": "20190731",
      "code_info": "Serial Numbers: 37701  35374  35937  35456  35327  35368  35432  35690  35301  35451  35553  35427  35743  35750  35762  35778  35788  35764  35736  35350  35369  35601  35664  35730  35783  35376  35714  35547  35325  35314  35522  35635  35660  35333  35511  35536  35560  35329  35655  35810  35335  35353  35359  35410  35599  35556  35361  35283  35828  35380  35893  35351  35815  35306  35390  35525  35921  35727  35441  35465  37702  37713  35182  35974  35704  37705  35269  35339  35345  35516  37709  35486  35431  35512  35739  35834  35535  35305  35326  35471  35472  35334  35577  35202  35279  35860  35600  35309  35684  37710  35663  35346  35789  37715  35649  35615  35665  35768  35945  35312  35423  35241  35253  35610  35291  35629  35836  35603  35916  35707  35728  35698  35402  35571  35583  35666  35550  35821  35406  35626  35240  35668  37700  35540  35518  35179  35249  35504  35562  35566  35567  35501  35502  35623  35490  35545  35476  35709  35712  35744  35232  35452  35695  35678  35266  35793  35338  35784"
    }
  ]
}