      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65413",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "345 E Main St",
      "address_2": "N/A",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) and the country of Japan.",
      "recall_number": "Z-2020-2013",
      "product_description": "00597202009\tArticular Surface\tCR ART SURF 12/PURPLE 09, Rx, Sterile;    00597202010\tArticular Surface\tCR ART SURF 12/PURPLE 10, Rx, Sterile;    00597202012\tArticular Surface\tCR ART SURF 12/PURPLE 12, Rx, Sterile;    00597202014\tArticular Surface\tCR ART SURF 12/PURPLE 14, Rx, Sterile;    00597202017\tArticular Surface\tCR ART SURF 12/PURPLE 17, Rx, Sterile;    00597202020\tArticular Surface\tCR ART SURF 12/PURPLE 20, Rx, Sterile;    00597202023                 Articular Surface\tCR ART SURF 12/PURPLE 23, Rx, Sterile;    Used in total knee arthroplasty and indicated for patients with severe pain.",
      "product_quantity": "192,355 all devices",
      "reason_for_recall": "Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues.  Zimmer is removing the d",
      "recall_initiation_date": "20130607",
      "center_classification_date": "20130822",
      "termination_date": "20141215",
      "report_date": "20130828",
      "code_info": "all codes"
    }
  ]
}