  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Malvern",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83203",
      "recalling_firm": "Siemens Medical Solutions USA, Inc",
      "address_1": "40 Liberty Blvd",
      "address_2": "N/A",
      "postal_code": "19355-1418",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution",
      "recall_number": "Z-2019-2019",
      "product_description": "AXIOM Artis dTA, Model Number 7008605",
      "product_quantity": "178",
      "reason_for_recall": "A tolerance issue in the power supply of the generator control above a specific value may cause generator A100 failure. As a result, the X-ray tube will not be supplied with the required voltage and X-ray will not be possible. The issue occurs sporadically; however, it may occur during an on-going procedure.",
      "recall_initiation_date": "20190612",
      "center_classification_date": "20190722",
      "termination_date": "20200918",
      "report_date": "20190731",
      "code_info": "Serial Numbers: 55163  55713  55304  55435  55567  55407  55412  55310  55470  55164  55187  55206  55091  55059  55204  55377  55307  55171  55368  55210  55133  55135  55477  55250  55318  55196  55230  55277  55423  55320  55387  55576  55438  55503  55046  55308  55139  55039  55270  55188  55314  55381  55550  55365  55574  55415  55109  55337  55322  55272  55157  55234  55548  55282  55575  55211  55410  55150  55079  55262  55259  55531  55388  55275  55276  55254  55569  55177  55269  55292  55041  55404  55425  55183  55441  55125  55369  55437  55136  55356  55191  55103  55400  55483  55083  55526  55088  55243  55056  55408  55715  55233  55416  55256  55074  55212  55153  55325  55312  55273  55213  55401  55170  55284  55469  55067  55048  55216  55241  55708  55202  55461  55154  55167  55342  55468  55219  55553  55217  55386  55396  55513  55279  55414  55347  55042  55294  55160  55286  55036  55362  55436  55375  55131  55062  55107  55411  55052  55096  55156  55289  55374  55705  55017  55223  55149  55141  55089  55305  55078  55227  55040  55200  55186  55184  55232  55358  55222  55313  55085  55090  55324  55045  55382  55350  55353  55405  55043  55439  55473  55070  55545  55434  55118  55455  55263  55274  55238"
    }
  ]
}