  "results": [
    {
      "status": "Ongoing",
      "city": "Telford",
      "state": "PA",
      "country": "United States",
      "classification": "Class I",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92443",
      "recalling_firm": "Draeger Medical, Inc.",
      "address_1": "3135 Quarry Rd",
      "address_2": "N/A",
      "postal_code": "18969-1042",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distributed Nationwide. OUS Distribution: Afghanistan, Algeria, Andorra, Angola, Antigua/Barbados, Argentina, Australia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Bolivia, Brunei Dar-es-S, Bulgaria, Burkina-Faso, Cambodia, Canada, Cayman Islands, Chile, China, Christmas Island, Columbia, Congo, Costa Rica, Croatia, Cuba, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Equatorial Guinea, Estonia, Ethiopia, Faeroe, Finland, France, French Polynesia, French Guiana, Gabon, Georgia, Germany, Ghana, Gibraltar, Greece, Guadeloupe, Guinea, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Ivory Coast, Japan, Jordan, Kazakhstan, Kenya, Kosovo, Kuwait, Laos,  Latvia, Lebanon, Libya, Liechtenstein, Lithuania, Macau, Malawi, Malaysia, Mali, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldavia, Mongolia, Morocco, Myanmar, Namibia, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, North Korea, Norway, Oman, Pakistan, Panama, Pap. New Guinea, Paraguay, Peru, Poland, Portugal, Qatar, Reunion, Romania, Russian Fed., Rwanda, San Marino, Saudi Arabia, Senegal, Serbia, Singapore, Sint Maarten, Slovenia, Slovakia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Suriname, Sweden, Switzerland, Taiwan, Thailand, Trinidad/Tobago, Tunisia, Turkey, Turkmenistan, Uganda, Ukraine, United Kingdom, Uruguay, Utd. Arab. Emirates, Uzbekistan, Vietnam, Yemen, Zambia, and Zimbabwe.",
      "recall_number": "Z-2018-2023",
      "product_description": "Oxylog 3000 Plus emergency and transport ventilator",
      "product_quantity": "300 units",
      "reason_for_recall": "Draeger has become aware of instances where the device stopped working when in use when transitioning from battery to AC operation.",
      "recall_initiation_date": "20230612",
      "center_classification_date": "20230707",
      "report_date": "20230719",
      "code_info": "Catalog No. 5704811 & 5704813; UDI-DI: 04048675398295; All Serial No.",
      "more_code_info": ""
    }
  ]
}