      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65413",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "345 E Main St",
      "address_2": "N/A",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) and the country of Japan.",
      "recall_number": "Z-2011-2013",
      "product_description": "00597201303\tFemur\tCR POR FM COMP SIZE CML Rx, Sterile;     00597201304\tFemur\tCR POR FM COMP SIZE CMR Rx, Sterile;     00597201403\tFemur\tCR POR FM COMP SIZE DML Rx, Sterile;     00597201404\tFemur\tCR POR FM COMP SIZE DMR Rx, Sterile;     00597201503\tFemur\tCR POR FM COMP SIZE EML Rx, Sterile;     00597201504\tFemur\tCR POR FM COMP SIZE EMR Rx, Sterile;     Used in total knee arthroplasty and indicated for patients with severe pain.",
      "product_quantity": "192,355 all devices",
      "reason_for_recall": "Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues.  Zimmer is removing the d",
      "recall_initiation_date": "20130607",
      "center_classification_date": "20130822",
      "termination_date": "20141215",
      "report_date": "20130828",
      "code_info": "all codes"
    }
  ]
}