      "status": "Terminated",
      "city": "Warsaw",
      "state": "IN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65413",
      "recalling_firm": "Zimmer, Inc.",
      "address_1": "345 E Main St",
      "address_2": "N/A",
      "postal_code": "46580-2746",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution-USA (nationwide) and the country of Japan.",
      "recall_number": "Z-2010-2013",
      "product_description": "00597001333\tFemur\tCR PRC SURF HDN FEM CO-NID CML Rx, Sterile;     00597001334\tFemur\tCR PRC SURF HDN FEM CO-NID CMR Rx, Sterile;     00597001433\tFemur\tCR PRC SURF HDN FEM CO-NID DML Rx, Sterile;     00597001434\tFemur\tCR PRC SURF HDN FEM CO-NID DMR Rx, Sterile;     00597001533\tFemur\tCR PRC SURF HDN FEM CO-NID EML Rx, Sterile;     00597001534\tFemur\tCR PRC SURF HDN FEM CO-NID EMR Rx, Sterile;     Used in total knee arthroplasty and indicated for patients with severe pain.",
      "product_quantity": "192,355 all devices",
      "reason_for_recall": "Surgeons are implanting components of the NexGen Cruciate Retaining (CR) Complete Knee System in combinations that are not approved (micro with standard). Includes CR Micro Femurs, Articular Surface, Patellae, instruments and provisionals combined with standard components. A March 2012 recall reiterated the proper combination labeling, but the mis match problem continues.  Zimmer is removing the d",
      "recall_initiation_date": "20130607",
      "center_classification_date": "20130822",
      "termination_date": "20141215",
      "report_date": "20130828",
      "code_info": "all codes"
    }
  ]
}