  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Minneapolis",
      "address_1": "17400 Medina Rd Ste 100",
      "reason_for_recall": "CareFusion identified that the EnVe Ventilator may not hold the set Positive End Expiratory Pressure (PEEP) value either intermittently or continuously or it may alarm for a Patient Circuit Fault.  The unit will activate either a Low PEEP or Patient Circuit Fault Alarm both audibly and Visually to alert the healthcare professional.  Ventilation delivery to the patient may be compromised.",
      "address_2": "",
      "product_quantity": "266",
      "code_info": "serial numbers:  26562, 43634, 43642, 43643, 44536, 44537, 44540, 44541, 44542, 44543, 44544, 44545, 44546, 44552, 44553, 44554, 44555, 44557, 44558, 44559, 44560, 44561, 44562, 44563, 44564, 44565, 45318, 45320, 45321, 45322, 45323, 45324, 45325, 45326, 45327, 45328, 45329, 45330, 45333, 45334, 45335, 45336, 45337, 45338, 45339, 45340, 45341, 45342, 45343, 45344, 45345, 45346, 45735, 45736, 45737, 45738, 45739, 45740, 45741, 45742, 45743, 45744, 45745, 45746, 45748, 45750, 45751, 45752, 45753, 45754, 45902, 45903, 45904, 45905, 45906, 45907, 45908, 45909, 45910, 45911, 46038, 46039, 46040, 46041, 46042, 46043, 46044, 46045, 46046, 46047, 46048, 46049, 46050, 46051, 46052, 46632, 46634, 46635, 46636, 46637, 46638, 46639, 46640, 46641, 46642, 46643, 46644, 46645, 47069, 47070, 47071, 47073, 47074, 47075, 47076, 47077, 47078, 47079, 47080, 47081, 47082, 47083, 47210, 47211, 47212, 47213, 47214, 47215, 47216, 47217, 47219, 47221, 47222, 47223, 47225, 47226, 47227, 47229, 47230, 47231, 47232, 47233, 47234, 47236, 47237, 47238, 47239, 47538, 47539, 47540, 47541, 47542, 47543, 47544, 47545, 47546, 47547, 47548, 47549, 47550, 47551, 47552, 47553, 47554, 47555, 47556, 47557, 47558, 47559, 47560, 47561, 47562, 47563, 47564, 47565, 47566, 47567, 48759, 48760, 48761, 48762, 48763, 48764, 48765, 48766, 48767, 48768, 48769, 48770, 48771, 48772, 48773, 48774, 48775, 48776, 48777, 48778, 48779, 48780, 48781, 48782, 48783, 48784, 48785, 48786, 48787, 48788, 48868, 48869, 48870, 48871, 48872, 48873, 48874, 48875, 48876, 48878, 48881, 48883, 48884, 48885, 48886, 48887, 48889, 48890, 48891, 48892, 48893, 48894, 48895, 48897, 50390, 50391, 50392, 50393, 50394, 50395, 50397, 50409, 50411, 50412, 50416, 50417, 50418, 50419, 50648, 50650, 50651, 50652, 50654, 50656, 50657, 50658, 50659, 50660, 50661, 50662, 50663, 50664, 50665, 50666, 50667, 50668, 50672, 51569, 51574",
      "center_classification_date": "20120716",
      "distribution_pattern": "Nationwide Distribution including AZ, CA, CO, FL, IL, KY, ME, MI, MO, MN, MT, ND, NJ, NM, NV, NY, OK, PA, SC, SD, TX , UT, WI, and WV.",
      "state": "MN",
      "product_description": "CareFusion EnVe Ventilator    Designed for use on patients who require respiratory support or mechanical ventilation and weigh a minimum of 5 kg (11 lbs). It is suitable for service in hospital and transport environments as a source of continuous or intermittent positive pressure ventilatory support, delivered invasively or non-invasively. It is not intended for homecare use.",
      "report_date": "20120725",
      "classification": "Class I",
      "openfda": {},
      "recalling_firm": "CareFusion 203, Inc.",
      "recall_number": "Z-2006-2012",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "62349",
      "termination_date": "20130211",
      "more_code_info": "",
      "recall_initiation_date": "20120619",
      "postal_code": "55447-1341",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}