      "status": "Terminated",
      "city": "Knoxville",
      "state": "TN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74192",
      "recalling_firm": "Siemens Medical Solutions USA Inc.",
      "address_1": "810 Innovation Dr",
      "address_2": "N/A",
      "postal_code": "37932-2562",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - US Nationwide",
      "recall_number": "Z-2005-2016",
      "product_description": "Biograph mCT-X 4R, MATERIAL NUMBER 10248670    The Siemens Biograph TruePoint systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.",
      "product_quantity": "83 units",
      "reason_for_recall": "Siemens Medical Solutions, Molecular Imaging has become aware of a potential for unexpected SUV values if a non-Siemens phantom is used for calibration.",
      "recall_initiation_date": "20160422",
      "center_classification_date": "20160621",
      "termination_date": "20170508",
      "report_date": "20160629",
      "code_info": "Serial Numbers: 1002,1003,1001,1004,1005,1006,1008,1009,1010,1007,1011,11013,11014,11015,11016,11017,11018,11020,11012,11019,11021,11022,11023,11026,11025,11024,11027,11044,11029,11028,11030,11031,11032,11033,11034,11037,11036,11038,11040,11042,11039,11035,11045,11046,11041,11047,11048,  11049,11050,11051,11052,11054,11055,11056,11043,11058,11059,11057,11061,11060,11062,11063,11064,11065,11066,11067,11068,11069,11070,21071,21073,21072,21074,31075,31076,31077,31083,31079,31080,31081,31082,31078,40001091760"
    }
  ]
}