  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Lebanon",
      "state": "TN",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92250",
      "recalling_firm": "Permobil",
      "address_1": "300 Duke Dr",
      "address_2": "N/A",
      "postal_code": "37090-8115",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide distribution.",
      "recall_number": "Z-2004-2023",
      "product_description": "PushTracker E2 component utilized with the SmartDrive MX2+ Power Assist Device (Part Numbers:  MX2-32K, MX2-32P, MX2-32PK, MX2-33D, MX2-33P, MX2-33PK, MX2-33S, MX2-150, MX2-167, MX2-32D); a wearable watch that communicates with a SmartDrive Power Assist  Device via the SmartDrive MX2+ Application",
      "product_quantity": "18,905 units",
      "reason_for_recall": "When multiple processes are running on the watch's Central Processing Unit (CPU), the application may fail unexpectedly. If this happens, the motor on the power assist device continues to run and the user may not be able to stop the device using tap gestures.",
      "recall_initiation_date": "20230510",
      "center_classification_date": "20230626",
      "report_date": "20230705",
      "code_info": "All units with software versions 1.0.00 to 1.0.04",
      "more_code_info": ""
    }
  ]
}