{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Lugano",
      "state": "N/A",
      "country": "Switzerland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85454",
      "recalling_firm": "Inpeco S.A.",
      "address_1": "Via San Gottardo 10",
      "address_2": "N/A",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide including in the states of Texas, West Virginia, South Carolina, Delaware, Louisiana, Illinois, Georgia, Florida, New York, Virginia, Arkansas, Wisconsin, California, Minnesota, Missouri, North Carolina, Utah, Michigan, Ohio, Alabama, Oklahoma, Tennessee, Kansas, Hawaii, Arizona, Washington",
      "recall_number": "Z-2003-2020",
      "product_description": "Inpeco FlexLab, ARCHITECT cSystem c8000/c16000 Interface Module, Part Numbers: FLX-208-00, FLX-208-10 - Product Usage: The FlexLab (aka Accelerator a3600) is a modular system designed to automate Pre-Analytical and Post-Analytical processing, sample handling in order to automate sample processing in the Laboratory. The systems consolidate multiple Analytical instruments into a unified workstation.",
      "product_quantity": "82",
      "reason_for_recall": "When analyzer sample carousel lid remains open or manual process is too long, Module timeout expiry errors could occur causing cross contamination of samples.",
      "recall_initiation_date": "20200407",
      "center_classification_date": "20200514",
      "termination_date": "20220728",
      "report_date": "20200520",
      "code_info": "Serial Numbers:  AML.001,  FLX.0104,  FLX.0105,  FLX.0106,  FLX.0107,  FLX.0108,  FLX.0112,  FLX.0114,  FLX.0115,  FLX.0116,  FLX.0118,  FLX.0119,  FLX.0120,  FLX.0124,  FLX.0125,  FLX.0126,  FLX.0131,  FLX.0133,  FLX.0134,  FLX.0135,  FLX.0136,  FLX.0140,  FLX.0141,  FLX.0142,  FLX.0143,  FLX.015,  FLX.0152,  FLX.0155,  FLX.0158,  FLX.0161,  FLX.0163,  FLX.0164,  FLX.017,  FLX.0175,  FLX.0178,  FLX.018,  FLX.0184,  FLX.0189,  FLX.019,  FLX.0196,  FLX.020,  FLX.021,  FLX.023,  FLX.026,  FLX.027,  FLX.030,  FLX.032,  FLX.033,  FLX.038,  FLX.040,  FLX.044,  FLX.045,  FLX.047,  FLX.051,  FLX.052,  FLX.053,  FLX.054,  FLX.055,  FLX.058,  FLX.059,  FLX.063,  FLX.064,  FLX.065,  FLX.066,  FLX.067,  FLX.068,  FLX.071,  FLX.072,  FLX.074,  FLX.075,  FLX.076,  FLX.077,  FLX.079,  FLX.080,  FLX.081,  FLX.091,  FLX.092,  FLX.095,  FLX.099,  FLX.101,  FLX.102,  FLX.103"
    }
  ]
}