  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Winterthur",
      "state": "",
      "country": "Switzerland",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "74206",
      "recalling_firm": "Zimmer Gmbh",
      "address_1": "Sulzer Allee 8",
      "address_2": "",
      "postal_code": "N/A",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Nationwide Distribution to AZ, CA, CO, GA, IL, KS, KY, MA, MD, MO, OH, PA, TN, and TX.",
      "recall_number": "Z-2003-2016",
      "product_description": "Dome Centric component (Reference 01.04227.005) from the Anatomical Shoulder (AS) System 2.0.  The Dome Centric connects the Humeral Stem to the Humeral Head and allows setting an inclination and retroversion positioning.  The Anatomical Shoulder System is intended for long-term implantation into the human shoulder joint in primary or revision, total or hemi shoulder arthroplasty. The system is intended to relieve pain and restore function in patients with adequate bone stock to support the prosthesis.",
      "product_quantity": "177",
      "reason_for_recall": "In some cases it has been difficult or not possible to disassemble the adjusted Dome Centric from the AS Humeral Rasp after initial positioning of the AS Humeral Trial Head which resulted in the whole construct of the Dome Centric with the still assembled AS Humeral Rasp taken out of the humeral canal. This could cause a delay in surgery time or the surgeon could decide to close the wound without finishing the surgery,  increasing the infection risk or second use of anesthesia necessary.",
      "recall_initiation_date": "20160517",
      "center_classification_date": "20160616",
      "termination_date": "20161031",
      "report_date": "20160622",
      "code_info": "Item Number: 01.04227.005  Lot Numbers:  2792457  2794578  2812029  2817212  2844463  2794576  2794579  2815118  2817213  2847568  2794577  2794580  2815119  2832862",
      "more_code_info": ""
    }
  ]
}