{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Knoxville",
      "address_1": "810 Innovation Dr",
      "reason_for_recall": "The bracket that is used to lift the cover has bolts that were not welded in potentially causing the cover to fall down during service.",
      "address_2": "",
      "product_quantity": "1 unit",
      "code_info": "Serial Number 21139",
      "center_classification_date": "20160615",
      "distribution_pattern": "Worldwide Distribution - US Nationwide and the countries of  China and Malaysia.",
      "state": "TN",
      "product_description": "BIOGRAPH mCT S(20) - 3R, MATERIAL NUMBER 10507786    Product Usage:  The Siemens Biograph mCT systems are combined X-Ray Computed Tomography (CT) and Positron Emission Tomography (PET) scanners that provide registration and fusion of high resolution physiologic and anatomic information.",
      "report_date": "20160622",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Siemens Medical Solutions USA Inc.",
      "recall_number": "Z-1997-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "74239",
      "termination_date": "20170201",
      "more_code_info": "",
      "recall_initiation_date": "20160425",
      "postal_code": "37932-2562",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}