  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-22",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Kent",
      "state": "WA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "68580",
      "recalling_firm": "Cenorin, LLC",
      "address_1": "6324 S 199th Pl Ste 107",
      "address_2": "",
      "postal_code": "98032-2135",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "distribution_pattern": "Worldwide Distribution - US (nationwide):  CA, DE, FL, MA, MI, NY, PA, TX, UT, WA, and WI and the countries of Uruguay and Taiwan.",
      "recall_number": "Z-1995-2014",
      "product_description": "HLD System, Model 610, Medical Device Cleaning and High Level Disinfection Washer/Pasteurizer.",
      "product_quantity": "25 HLD systems",
      "reason_for_recall": "The temperature sensor/control system in the HLD Systems Model 610 may provide a false temperature reading. This caused the water to remain at ambient temperature throughout the pasteurization cycle rather than using hot water.",
      "recall_initiation_date": "20130124",
      "center_classification_date": "20140706",
      "termination_date": "20150210",
      "report_date": "20140716",
      "code_info": "610601, 610603, 610703, 610704, 610705, 610707, 610708, 610802, 610803, 610804, 610901, 610903, 610904, 610905, 611001, 611002, 611004, 611101, 611102, 611103, 611201, 611202, 611203, 610902L, 611204.",
      "more_code_info": ""
    }
  ]
}