  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-29",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Rochester",
      "state": "NY",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "65866",
      "recalling_firm": "Getinge USA Inc",
      "address_1": "1777 E Henrietta Rd",
      "address_2": "N/A",
      "postal_code": "14623-3133",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution -  USA and the country of Canada.",
      "recall_number": "Z-1994-2013",
      "product_description": "Getinge Assure Accufast Biological Indicators, Catalog Number 61301606636 (50 Pack) and 61301606637 (100 Pack), STEAM ---  Manufactured for: Getinge USA, Inc. 1777 E. Henrietta Rd., Rochester, NY 14623.     Getinge Assure AccuFast Biological Indicators (BIs) are designed for biological testing of 121¿C, 132¿C, 134¿C and 135¿C steam sterilization cycles. The AccuFast BI is a self-contained biological indicator containing Geobacillus stearothermophilus inoculated on a paper carrier and placed in a plastic vial with a small glass ampule containing sterile culture medium. After activation, the plastic vial serves as a culture tube. The chemical indicator on the label changes from blue to black when exposed to steam, distinguishing exposed from unexposed units. This indicator does not indicate adequate sterilization.",
      "product_quantity": "Catalog # 61301606636: 78 Packs (3900 BIs); Catalog # 61301606637: 43 Packs (4300 BIs)",
      "reason_for_recall": "AccuFast Biological Indicator Lot SR-430 may have been exposed to conditions affecting its viability. This could result in a false negative (or pass) when exposed to sub-lethal sterilization parameters.",
      "recall_initiation_date": "20130802",
      "center_classification_date": "20130821",
      "termination_date": "20170223",
      "report_date": "20130828",
      "code_info": "LOT SR-430 (Expiration 02/06/2014)"
    }
  ]
}