  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "country": "United States",
      "city": "Neptune",
      "address_1": "1933 Heck Ave",
      "reason_for_recall": "Certain lots of Heparin Lock Flush, USP Syringes (5 ml Fill in 10 ml Syringe) failed Anti-Factor IIa assay for potency during testing of retention samples for heparin potency.",
      "address_2": "",
      "product_quantity": "184,500",
      "code_info": "Product Code, Lot number, Expiration date.    513604 3123326 7/1/2016  E0500-55 3123349 7/1/2016  513602 3126577 8/1/2017",
      "center_classification_date": "20160614",
      "distribution_pattern": "US Nationwide distribution.",
      "state": "NJ",
      "product_description": "Excelsior Disposable Syringe Heparin Lock Flush, USP syringe (5 mL Fill in 10 mL syringe).  For flushing of IV catheters and IV tubing only.",
      "report_date": "20160622",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "Excelsior Medical Corp",
      "recall_number": "Z-1989-2016",
      "initial_firm_notification": "Letter",
      "product_type": "Devices",
      "event_id": "74045",
      "termination_date": "20170808",
      "more_code_info": "",
      "recall_initiation_date": "20160428",
      "postal_code": "07753-4428",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}