      "recalling_firm": "Alliance Partners LLC",
      "address_1": "121 Interpark Blvd Ste 601",
      "address_2": "N/A",
      "postal_code": "78216-1842",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide Distribution in the states LA, TX, and PR.",
      "recall_number": "Z-1989-2015",
      "product_description": "Nakoma-SL Anterior Cervical Plating System's Variable Single Barrel Drill Guide    Product Usage:  The Nakoma-SL ACP System is intended for anterior intervertebral screw fixation of the cervical spine at levels C2-T1. The Nakoma ACP System is indicated for temporary stabilization of the anterior spine during the development of cervical spine fusion in patients with the following indications: Degenerative Disk Disease, Trauma, Tumors, Deformities or curvatures, Psuedoarthrosis, Failed previous fusions, Spondylolistesis, and Spinal Stenosis.",
      "product_quantity": "21 units",
      "reason_for_recall": "Inadequate interference between the screw and the plate when screw implanted into the plate at angulation greater than 7 degrees can result in the screw advancing through the plate during implantation.",
      "recall_initiation_date": "20150609",
      "center_classification_date": "20150707",
      "termination_date": "20151009",
      "report_date": "20150715",
      "code_info": "Lot numbers SM59161 and SM60278"
    }
  ]
}