{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Mansfield",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "85629",
      "recalling_firm": "Covidien Llc",
      "address_1": "15 Hampshire St",
      "address_2": "N/A",
      "postal_code": "02048-1113",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-1988-2020",
      "product_description": "Dover 100% Silicone Premium Urine Meter Foley Tray, 400mL, 14 Fr/Ch (4.7 mm), 5 mL Catheter Pre-connected to 2000 mL Drainage Bag with Luer-Lock Sampling, Temperature Sensor and Adhesive Catheter Securement,  Sterile, Product Code: P4P14TSD - Product Usage: for Drainage of the Urinary bladder and simultaneous monitoring of temperature. Prescription use only.",
      "product_quantity": "3470 units",
      "reason_for_recall": "Temperature sensing catheters failed to meet the  established calibrated temperature sensing tolerance of +/- 0.2¿C. Failure to correctly measure patient  body temperatures could lead to an incorrect diagnosis and insufficient treatment",
      "recall_initiation_date": "20200430",
      "center_classification_date": "20200514",
      "termination_date": "20231214",
      "report_date": "20200520",
      "code_info": "Lot Numbers: 1909308464, 1933000164"
    }
  ]
}