{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Irving",
      "state": "TX",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "83189",
      "recalling_firm": "Abbott Laboratories, Inc",
      "address_1": "1921 Hurd Dr",
      "address_2": "N/A",
      "postal_code": "75038-4313",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Distribution was nationwide, including Puerto Rico.  There was government/military distribution.    Foreign distribution was made to Canada, Albania, Algeria, Andorra, Argentina, Armenia, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovi, Botswana, Brazil, Brunei, Bulgaria, Chile, China, Costa Rica, Croatia, Curacao (Netherlands Antilles), Cyprus, Czech Republic, Democratic Republic of the Congo, Denmark, Ecuador, Egypt, Estonia, Finland, France, Gaza & Jericho, Germany, Greece, Greenland, Honduras, Hong Kong, India, Iraq, Ireland, Israel, Italy, Ivory Coast, Jamaica, Jordan, Kazakhstan, Kenya, Kuwait, Kyrgyzstan, Latvia, Lebanon, Libya. Lithuania, Luxembourg, Macedonia, Malaysia, Maldives, Mali, Mauritania, Mexico, Moldova, Montenegro, Morocco, Namibia, Netherlands, New Zealand, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Republic of Yemen, Romania, Russia, Rwanda, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Tanzania, Thailand, Trinidad & Tobago, Tunisia, Turkey, Turks and Caicos, Tunisia, UAE, United Kingdom, Uruguay, and Viet Nam.",
      "recall_number": "Z-1988-2019",
      "product_description": "Immunoglobulin A is an in vitro diagnostic assay for the quantitative determination of  immunoglobulin A in human serum or plasma. Antigen in the sample bonds to the  specific antibody in the reagent, forming an immune complex. The immune complex  causes an increase in light scattering, measured by reading turbidity at 700 nm, which correlates with the concentration of immunoglobulin A in the sample.",
      "product_quantity": "12,995 kits",
      "reason_for_recall": "An update to the Reagents and Specimen Collection and Handling/Preparation for Analysis sections of the IFU was required due to different issues with the various products.",
      "recall_initiation_date": "20180516",
      "center_classification_date": "20190719",
      "termination_date": "20210816",
      "report_date": "20190731",
      "code_info": "All IFU's"
    }
  ]
}