{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-03-18",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Center Valley",
      "state": "PA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "92325",
      "recalling_firm": "Olympus Corporation of the Americas",
      "address_1": "3500 Corporate Pkwy",
      "address_2": "",
      "postal_code": "18034-8229",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "US Nationwide distribution.",
      "recall_number": "Z-1987-2023",
      "product_description": "Single Use Repositionable Clip QUICKCLIP PRO 165 CM 10/BX- Intended for hemostatic metal clip for the GI Tract  Catalog Number: HX-202LR.A",
      "product_quantity": "340 pieces",
      "reason_for_recall": "Complaints on clip deployment occurring during clinical procedures, leading to hazardous situations where clips fail to achieve their expected function or deploy incorrectly, potentially causing harms to patients. The complaints include reports that: 1) the clip arm does not open when the user pushes the slider, 2) the clip arm does not close when the user pulls the slider and 3) the clip detaches from the target tissue earlier than expected, after being deployed in a procedure.",
      "recall_initiation_date": "20230504",
      "center_classification_date": "20230621",
      "report_date": "20230628",
      "code_info": "UDI-DI: 04953170385919 Lot Numbers: 24K, 27K, 29K, 2XK, 2YK",
      "more_code_info": ""
    }
  ]
}