  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-15",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Terminated",
      "city": "Hanover",
      "state": "MD",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "71443",
      "recalling_firm": "BTE Technologies, Inc.",
      "address_1": "7455 New Ridge Rd Ste L",
      "address_2": "N/A",
      "postal_code": "21076-3143",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide Distribution - Nationwide Distribution and to the countries of : Austria, Belgium, Canada, France, Germany, Greece, Hong Kong, Italy, India, Japan, South Korea, Mexico, P.R. China, Poland, Republic of South Africa, Romania, Saudi Arabia, Singapore, Spain, Switzerland, Thailand, and United Kingdom.",
      "recall_number": "Z-1986-2015",
      "product_description": "Primus (THE BTE WS30), model PrimusRS (PRRS) and Primus+ (PR30) a measuring exercise equipment device designed to provide active and passive torque resistance to patient motion during rehabilitation therapy.  The Primus that includes the optional Chop / Lift Bar.",
      "product_quantity": "510 Devices",
      "reason_for_recall": "On the Primus that includes the optional Chop / Lift Bar, there is a remote possibility that the Chop / Lift Bar could get accidently detached from the snap hook in specific use scenarios when the bar is rotated and twisted around the snap hook.  The snap hook could open, resulting in possible injury to the user due to a fall.",
      "recall_initiation_date": "20150409",
      "center_classification_date": "20150706",
      "termination_date": "20160510",
      "report_date": "20150715",
      "code_info": "Primus (THE BTE WS30), model PrimusRS (PRRS) and Primus+ (PR30)"
    }
  ]
}