{
  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-07-08",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Sparks",
      "state": "MD",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "98657",
      "recalling_firm": "Becton Dickinson & Co.",
      "address_1": "7 Loveton Cir",
      "address_2": "",
      "postal_code": "21152-9212",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the country of Hong Kong.",
      "recall_number": "Z-1983-2026",
      "product_description": "BD GasPak EZ Campy Pouch System. Catalog Number: 260685. for microbiologic use",
      "product_quantity": "2,200 units",
      "reason_for_recall": "BD identified an issue affecting certain lots of BD GasPak EZ CO2 and BD GasPak EZ Campy Gas Generating Pouch Systems, where gas generating sachets may produce carbon dioxide levels below specification.",
      "recall_initiation_date": "20260407",
      "center_classification_date": "20260429",
      "report_date": "20260506",
      "code_info": "Catalog Number: 260685; UDI-DI: 30382902606858; Lot Number: 5329863.",
      "more_code_info": ""
    }
  ]
}