  "meta": {
    "disclaimer": "Do not rely on openFDA to make decisions regarding medical care. While we make every effort to ensure that data is accurate, you should assume all results are unvalidated. We may limit or otherwise restrict your access to the API in line with our Terms of Service.",
    "terms": "https://open.fda.gov/terms/",
    "license": "https://open.fda.gov/license/",
    "last_updated": "2026-04-01",
    "results": {
      "skip": 0,
      "limit": 1,
      "total": 1
    }
  },
  "results": [
    {
      "status": "Ongoing",
      "city": "Cambridge",
      "state": "MA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "87974",
      "recalling_firm": "Philips North America Llc",
      "address_1": "222 Jacobs St",
      "address_2": "N/A",
      "postal_code": "02141-2289",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "Letter",
      "distribution_pattern": "Worldwide distribution - US Nationwide and the countries of Canada, Mexico, AE,  AR, AT, AU, BD, BE, BR, CH, CL, CN, CO, CU, CY, CZ, DE, DK, DZ, EE, EG, ES, FI, FR, GB, GE, GI, GL, GR, HK, HU, ID, IE, IL, IN, IQ, IR, IT, JO, JP, KE, KG, KH, KR, LA, LB, LK, LT, LY, MM, MN, MU, MV, MY, NI, NL, NO, NZ, OM, PA, PF, PH, PK, PL, PS, PT, QA, RE, RU, RW, SA, SE, SG , SI, SK, TH, TN, TR, UZ, VN, XK, YE, ZA, ZW.",
      "recall_number": "Z-1983-2021",
      "product_description": "Continuous CT Fluoroscopy DVI Ceiling Mount   CCT Arm DVI Option - Monitor Bracket Assembly for use with CT and PET/CT systems.     BB (Big Bore) Oncology, BB (Big Bore) Radiology:    728242/728243/728244/728272 - Brilliance Big Bore",
      "product_quantity": "1,301 Monitor Bracket Assemblies total",
      "reason_for_recall": "Significant wear of the Monitor Bracket Assembly shaft  goes unrecognized could potentially lead to the Monitor Bracket Assembly detaching from the Suspension Arm and result in potential hazard for the monitor to fall while remaining attached to the  cables resulting in patient, operator, bystander, or service person  to serious injury.",
      "recall_initiation_date": "20210501",
      "center_classification_date": "20210630",
      "report_date": "20210707",
      "code_info": "Model/ Serial Number/UDI: N/A  728242 20004 N/A  728242 768007 (01)00884838095168  728242 20013 N/A    728243 75107 N/A  728243 7729 N/A  728243 7786 N/A  728243 7741 N/A  728243 7820 N/A  728243 7892 N/A  728243 75027 N/A  728243 75029 N/A  728243 7941 N/A  728243 75047 N/A  728243 7467 N/A  728243 7831 N/A  728243 7831 N/A  728243 7560 N/A  728243 75067 N/A  728243 7689 N/A  728243 7442 N/A  728243 75111 N/A  728243 7927 N/A  728243 7558 N/A  728243 7769 N/A  728243 7284 N/A  728243 75066 N/A  728243 7905 N/A  728243 7800 N/A  728243 7724 N/A  728243 7946 N/A  728243 7385 N/A  728243 7192 N/A  728243 7211 N/A  728243 7891 N/A  728243 75041 N/A  728243 7269 N/A  728243 7064 N/A  728243 7116 N/A  728243 7214 N/A  728243 7795 N/A  728243 7701 N/A  728243 7810 N/A  728243 7743 N/A  728243 75064 N/A  728243 75120 N/A  728243 750005 N/A  728243 750013 N/A    728244 760014 (01)00884838059450(21)760014  728244 760013 (01)00884838059450(21)760013  728244 76230 N/A  728244 76259 (01)00884838059450(21)76259  728244 76238 (01)00884838059450(21)76238  728244 76147 (01)00884838059450(21)76147  728244 76178 (01)00884838059450(21)76178  728244 7764 N/A  728244 760045 (01)00884838059450(21)760045  728244 760023 (01)00884838059450(21)760023  728244 75123 N/A  728244 760047 (01)00884838059450(21)760047  728244 76109 (01)00884838059450(21)76109  728244 760056 (01)00884838059450(21)760056  728244 7653 N/A  728244 76249 (01)00884838059450(21)76249  728244 76228 (01)00884838059450(21)76228  728244 76299 (01)00884838059450(21)76299  728244 7926 N/A  728244 7533 N/A  728244 7917 N/A  728244 76095 (01)00884838059450(21)76095  728244 7874 N/A  728244 7452 N/A  728244 76113 (01)00884838059450(21)76113  728244 76140 (01)00884838059450(21)76140  728244 76078 (01)00884838059450(21)76078  728244 7781 N/A  728244 76079 (01)00884838059450(21)76079  728244 760015 (01)00884838059450(21)760015  728244 760029 N/A  728244 760011 (01)00884838059450(21)760011  728244 7694 N/A  728244 76248 N/A  728244 7747 N/A  728244 7773 (01)00884838059450(21)7773  728244 7826 N/A  728244 76290 (01)00884838059450(21)76290  728244 7915 N/A  728244 7721 N/A  728244 76268 (01)00884838059450(21)76268  728244 76177 (01)00884838059450(21)76177  728244 760003 (01)00884838059450(21)760003  728244 760004 (01)00884838059450(21)760004  728244 76214 (01)00884838059450(21)76214  728244 76250 (01)00884838059450(21)76250"
    }
  ]
}