      "country": "United States",
      "city": "Palm Beach Gardens",
      "address_1": "4500 Riverside Drive",
      "reason_for_recall": "Several attachments and handpieces failed to meet the expected sterility assurance level (SAL) when sterilized by the method that is specified in the current Directions for Use (DFU) provided with the device.",
      "address_2": "",
      "product_quantity": "115,988 units",
      "code_info": "DFU:    18-0047: Revisions A & B   18-0074: Revision A; and    18-0105: Revisions A.    Operating Manuals:   19-0010: Revisions A, and   19-0013: Revisions A, B, & C.",
      "center_classification_date": "20140703",
      "distribution_pattern": "Worldwide Distribution -- USA, including the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD,  ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX,UT, VA, VT, WA, WI, WV, WY, and Hawaii, and the territories of Puerto Rico, US Virgin Island and Guam; and, the countries of Argentina, Australia, Austria, Belgium, Brazil, Cameroon, Canada,  Chile, Costa Rica, Czech Republic, Dominican Republic, Egypt, Finland, France, Germany, Grand Cayman, Hong Kong, India, Israel, Jamaica, Japan, Korea, Lithuania, Malaysia, Mexico, Nassau, Netherlands, Nigeria, Palestine, Panama, Poland, Portugal, Saudi Arabia, Singapore, South Africa, Spain, Sweden, Switzerland, Syria, Thailand, Turkey, United Arab Emirates, United Kingdom, and Venezuela.",
      "state": "FL",
      "product_description": "Anspach Pneumatic and Electric Motor Systems.    Intended for cutting and shaping bone, including the spine and cranium.",
      "report_date": "20140709",
      "classification": "Class II",
      "openfda": {},
      "recalling_firm": "The Anspach Effort, Inc.",
      "recall_number": "Z-1983-2014",
      "initial_firm_notification": "Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit",
      "product_type": "Devices",
      "event_id": "68496",
      "termination_date": "20160129",
      "more_code_info": "",
      "recall_initiation_date": "20140508",
      "postal_code": "33410-4235",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "status": "Terminated"
    }
  ]
}