  "results": [
    {
      "status": "Terminated",
      "city": "Brea",
      "state": "CA",
      "country": "United States",
      "classification": "Class II",
      "openfda": {},
      "product_type": "Devices",
      "event_id": "82919",
      "recalling_firm": "Beckman Coulter Inc.",
      "address_1": "250 S Kraemer Blvd",
      "address_2": "N/A",
      "postal_code": "92821-6232",
      "voluntary_mandated": "Voluntary: Firm initiated",
      "initial_firm_notification": "E-Mail",
      "distribution_pattern": "US: CA and NY  OUS: United Kingdom, Switzerland and Germany",
      "recall_number": "Z-1982-2019",
      "product_description": "Kaluza C Flow Cytometry Software, RX Only in the U.S.A., Beckman Coulter, Inc. for the following  software devices:    (A) Kaluza C Perpetual License  (B) Kaluza C Single Use 1 Year License  (C) Kaluza C 5 User Network  (D) Kaluza C 10 User Network  (E) Kaluza C Perpetual Educational License  (F) Kaluza C Single User 1 Year educational License  (G) Kaluza C 10 User Network Educational License  (H) Kaluza C 5 User Network Educational License  (I) Kaluza C Flow Cytometry Software",
      "product_quantity": "41 units",
      "reason_for_recall": "Compatibility issue between the software and Microsoft updates to Windows 7, 8, and 10.  The compatibility issue causes the software to be unusable  which could result in a delay of reporting results.",
      "recall_initiation_date": "20180917",
      "center_classification_date": "20190717",
      "report_date": "20190724",
      "code_info": "A. C10574/UDI: 15099590705084  B. C10575/UDI: 15099590705091  C. C10576/UDI: 15099590705114  D. C10577/UDI: 15099590705107  E. C10578/UDI: 15099590705121  F. C10579/UDI: 15099590705138  G. C10580/UDI: 15099590705145  H. C10581/UDI: 15099590705152  I.  C21166/UDI: 15099590705169"
    }
  ]
}